Super Static Field Meter Scale
(With built-in one-chip microcomputer)
VESSEL’s EYE-01 is a palm-sized electrostatic measuring device ideal for measuring polarity and intensity of static charge on parts and assembly lines. Also, it’s very useful for checking ion balance performance of various ionizers.
What makes the EYE-01 so user-friendly and easy to use?
Easy-to-read Wide LCD screen display- Upper left graph display – indicates negative polarity
- Upper right graph display – indicates positive polarity
- Lower section display – shows numerical values, voltage measurement unit, and functions in use.
- Battery icon indicates when to replace the battery
- LED lighting function to read measurement values in the dark
- “Static charge measurement mode (Standard mode)” checks if the object is statically charged. It’s calibrated for voltage range 0 to 20kV.
- ± 0 ± 1.49kV (Low range)
- ± 1.0 ± 20.0kV (High range)
Ion balance measuring mode for checking performance on ionizers
- Compact design and lightweight makes it easy for operator to perform periodic verification of ionizers, including units that are installed in hard-to-reach places like in automated machinery or workbenches.
- Can be used to measure low voltages up to 200V in the ion balance mode.
- Ion balance plate included (charge plate).
- A measuring distance of 0.98” ± 0.02” (25mm±0.5mm) can be easily checked by "Focus Rings" beamed from the two red LEDs.
- Continuous mode - allows operator to take readings longer than 5 minutes. Simply turn on the EYE-01 without activating the “auto power off function” by keeping the power button pressed in for 3 seconds.
- Hold mode - enables measurements in hard-to-reach areas. When taking measurements in space-restricted areas, simply press the “hold mode” and the measurement value will be kept on the display screen.
- Digital zero mode - allows operator to adjust the reading to zero. Simply press “reset” to adjust the displayed reading and bar graph to the neutral state.
- Ion balance mode - takes measurement readings to check the ion balance performance on various ionizers.
- EYE-01 is calibrated by testing equipment that’s proved to be fully compliant with ISO standards. Inspection (calibration) certificate can be issued for additional fee.
- CE RoHS compliant
Static Voltage Measuring Range | ± 0 ~ ± 1.49 kV (Low range) ± 1.0 ~ ± 20.0 kV (High range) |
Ion Balance Voltage Measuring Range | 0 ~ ± 200v |
Measuring Distance | 0.98" ± 0.02" (25mm ± 0.5mm) (between charged object and field meter) |
Accuracy | ± 10% |
Ambient Conditions | 50ºF ~ 104ºF (10ºC ~ 40ºC), 60% RH or lower |
Display | Large LCD display (Digital and bar graph) |
Power | 9V. 6F22Y manganese battery: 1 PCE (Life: Approx. 30 hrs) |
Weight w/Battery | Approx. 0.31 lbs (140g) without ion balance plate [Ion balance plate alone: Approx. 0.07 lbs. (30g)] |
Accessories | Charge plate for ion balance measurement Ground cord (w/alligator clip), Battery (006P), soft case |
Applications | For checking the performance of ionizers, locating and measuring electrostatic charge build-up, deciding best location for installing ionizers and testing their efficiency. |
For Video of Vessel Eye-01, pls click
------------------------------
Notes:
1. cross reference from other manufacturers:
a. 3M 718 + 3M™ Model Air Ionizer Test Kit 718A
b. Simco FMX-002 , Simco Fmx-003
Notes sa paggamit:
1. It is a must to Pls read the Catalog & the Instruction Manual first before gamitin ang Eye-01. it is to familiarize your self sa parts, but very easy lang ang paggamit ng Vessel eye-01.
2. put the battery
3. put the grounding cord dun sa slot sa right side nya. ( recommended). Recommended na iground siya, but if pag ang application is
hindi abot yung grounding cord, it is ok kahit hindi nakaground, but more
accurate ang reading pag nakaground.
If naka anti-static wrist strap yung users, or naka conductive shoes or heel grounder siya, or if nasa EPA ( ESD protected Area) yung operators, then hindi na need na iground siya kasi yung main body ng Vessel eye-01 is made from conductive resin
4. I power ON
5. if you will use it to measure the static charges, then tanggalin ang
charge plate na nasa top head para lumitaw yung sensor, matatanggal mo yung charge plate by pushing the side buttons. Ilagay sa ilalim yung charge plate sa pamamagitan ng pagsuksok sa slot
6. then you can measure the statics, the correct distance is
pag nagcoverge na yung 2 laser circle
7. pag tapos na ang reading, i Off ang eye01, then pls bring
back yung charge plate sa original position nya nya sa Top position, this will
also protect the sensor
8. If you will use it to measure the Ionizer performance- Ion Balance, then I power ON lang ang unit. the charge plate is in the top position, then push the ION Balance button, then it will read the Ion balance of the Ionizer. The Ionizers are in good working condition if the reading is 0 to 50V. mas malapit sa zero mas better ang quality ng ionizers.